Test Categories
A Test Category is a list of practical tests (or Test Types) to be performed on equipment, and the associated Upper and Lower Limits. Former WinPATS™ users may have known them as Equipment Types.
The test categories page consists of group tests and individual tests. When creating new pieces of equipment you can select from either of these. There are many test categories however there are only a few that will be used frequently. The most frequently use categories are listed in the first section.
Commonly Used Test Categories:
In Built Test Categories
Manual Trigger on TnT+
Description
TNT_CL1
F1 Key
Class one, Earth bond & Insulation Test @500Vdc (default)
TNT_CL1_SRG
F1 key
Class one, Earth bond & Insulation Test @250Vdc
TNT_CL2
F2 Key
Class two, Insulation Test only @500Vdc (default)
TNT_CL2_SRG
F2 key
Class two, Insulation Test only @250Vdc
TNT_EXTLEAD
F3 Key
Extension lead test, Earth bond, Insulation @500V & Polarity tests
TNT_EXTLEAD_SRG
F3 key
Extension lead test, Earth bond, Insulation @250V & Polarity tests
TNT_LEAKAGE
Menu 2, F2 Key
See leakage section below. For El, Plus and 3PL models only.
TNT_RCD
RCD Menu, F1 Key
See RCD section below. For Plus and 3PL models only.
Leakage Test Categories:
Caution!! All leakage tests apply power to DUT.
In Built Test Categories
Manual Trigger on TnT+
Description
TNT_LEAKAGE_CL2EXT
Menu 2, F2 Key
For class 2 devices. Pass level of < 1mA.
TNT_LEAKAGE_RCD
Menu 2, F2 Key
If appliance has RCD mains plug installed. Performs an earth test then a leakage test. Pass level of < 2.5mA.
TNT_LEAKAGE_CL1
Menu 2, F2 Key
For class 1 devices. Performs an earth test then a leakage test. Pass level of < 5.0mA.
RCD Test Categories:
In Built Test Categories
Manual Trigger on TnT+
Description
TNT_RCD_TRIP_TM
RCD Menu, F1 key
RCD trip time test based on the RCD settings in the WinPATS options (file menu ---> options).
TNT_RCD_TRIP_TM_10
RCD Menu, F1 key
Preset 10mA test, RCD trip time test for 10mA RCD devices.
TNT_RCD_TRIP_TM_30
RCD Menu, F1 key
Preset 30mA test, RCD trip time test for 30mA RCD devices.
TNT_RCD_CUR_RMP
RCD Menu, F2 key
RCD current ramp test. Increased RCD leakage current till trip current level is found.
Other Test Categories:
In Built Test Categories
Manual Trigger on TnT
Description
TNT_POWER_TEST
Menu 2, F1 Key (TnT+ Only)
Caution!! Applies power to DUT. Measures Volts, current, PF, Watts, VA, of DUT whilst in operation
TNT_LOG_POW_TEST
Computer Trigger only
Caution!! Applies power to DUT. Performs a continuous power test with the result logged by the computer. Power log test setting defined in WinPATS options menu.
TNT_CL1_eIM
F1 key
Class 1 test to medical standards. Not applicable if TnT is not designed for testing standards.
TNT_CL2_eIM
F2 Key
Class 2 test to medical standards. Not applicable if TnT is not designed for testing standards.
TNT_CL1_SRG_eIM
F1 key
Class 1 @ 250V insulation test & tested to medical standards. Not applicable if TnT is not designed for testing standards.
TNT_LEAKAGE_eIM
Menu 2, F2 key
Leakage test to medical standards. Not applicable if TnT is not designed for testing standards.
Test Categories (Single Test):
Most test categories used are a combination of tests. For example, a class 1 test consists of an earth bond and insulation test. If you want to use the individual tests for your test category then choose one from the list below.
In Built Test Categories
Manual Trigger on TnT
Description
TNT_VISUAL
Computer Trigger only
Brings up the visual check box when testing appliance. Usefull for asset management.
TNT_EARTH
Computer Trigger only
Earth bond test.
TNT_INS250
F2 key
250V insulation test
TNT_INS500
F2 key
500V insulation test
TNT_POLARITY
Computer Trigger only
polarity check
TNT_EARTH_eIM
Computer Trigger only
Earth bond test to medical standards. Not applicable if TnT is not designed for testing standards.
TNT_LEAKAGE
Computer Trigger only
Performs a leakage test with a pass fail <5mA.
3PL Test Categories - Only for 3PL TnT Models:
3PL TnT models can also select the previously listed test categories to do single phase testing.
In Built Test Categories
Manual Trigger on TnT
Description
TNT3P_CL1
Menu 3, F1 key
3 phase class 1 test. Earth bond & Insulation Test @500Vdc (default)
TNT3P_CL1_SRG
Menu 3, F1 key
3 phase class 1 test. Earth bond & Insulation Test @250Vdc
TNT3P_EXTLEAD
Menu 3, F2 key
3 phase extension lead testing. Earth bond, Insulation @500V & Polarity tests
TNT3P_EXT_SRG
Menu 3, F2 key
3 phase extension lead testing. Earth bond, Insulation @250V & Polarity tests
TNT3P_4PIN_LKGE
Menu 3, F3 key
4 Pin appliance, Leakage test with the _CL1, _CL2 & _RCD test category type. See leakage test categories for details.
TNT3P_5PIN_LKGE
Menu 3, F3 key
5 Pin appliance, Leakage test with the _CL1, _CL2 & _RCD test category type. See leakage test categories for details.
TNT3P_RCD1_CUR_RMP
RCD Menu, F2 key
RCD current ramp testing with _RCD1, _RCD2 & _RCD3 phase selection.
TNT3P_RCD1_TRIP_TM
RCD Menu, F1 key
RCD trip time testing with _RCD1, _RCD2 & _RCD3 phase selection.
Adding a Test Category
Note: that new test categories can be created however they may not function with the appliance tester. This is because the TnT® G4-Series appliance tester needs pre-defined test procedures. If you make custom ones then the unit may not understand.
It is ok to modify the upper and lower limits of the test categories or any other variable. Doing this will still allow the test to be executed on the TnT® G4-Series tester.
- Click
on the main screen toolbar - Click

- The Test Category Details window will appear
- Enter a unique name for the new Test Category
- View the Available Test Types on the right and click the relevant tests to assign. By holding down the Ctrl key, you can select multiple tests. Selected Test Types will turn dark blue.
If a Test Type you require is not available, or the name is correct but the unit of measure is wrong, you will need to add more Test Types
- Click

- The Test Categories window will appear
- Enter a name for the new test type (e.g. Visual)
- Select the appropriate UOM (Unit of Measure) from the right column
If you do not see the UOM you want in the list
- Enter a new one
- Click Add to list
- Repeat this process for as many Test Types as you wish. Several Test Types can have the same name but their UOM must differ
If you make a mistake
- Select the Test Type from the list, click Remove
- Click OK when all new Test Types have been added
- When you return, the new Test Type will be automatically assigned to the new Test Category
To assign the Test Types selected on the right
- Click
to assign. Double clicking will also assign - The order of the Assigned Test Types is also the order that tests will be performed in
To change the order of test types
- Select single or multiple Test Type/s
- Click the Move Up
or Move Down
buttons as needed - Click OK when done
- The new Test Category will appear with the other Test Categories and can be assigned to equipment
Converting Old Test Categories
If you are going to use a TnT(+) to test equipment when you previously haven't, you may need to migrate to TnT Test Categories. For example, you used to use an Equipment Type called CLASS1 but now you will be using a TnT(+) for testing.
To Convert a Test Category
- Click Test Categories on the main screen toolbar
- Double click the old Test Category (e.g. CLASS1)
- The Test Category Details screen will appear
- Tick the Migrate to Another box and then select the TnT Test Category (e.g. TNT_CL1)
- Click OK
- Click Yes to the confirmation message
- All equipment formerly tested otherwise will now be tested by TnT(+)
To delete the old Test Category, if desired
- Simply select the Test Category (e.g. CLASS1)
- Click Delete on the main screen toolbar or right click the old Test Category and choose Delete
The test categories provide by this program are designed specifically for the TnT® G4-Series units only. Additional test categories can be created but they can not be control types for test equipment. Test Created type codes can be developed for inspection programs and checking procedures when used with PDA’s or PC/Laptops. Eg. Exit Lighting tests, Anesthetic Equipment maintenance and inspection.
- Click